Mode of operation-Wirkungsweise-Fonctionnement-Funzionamento-Funcionamiento-Werking-Funktion-Toimintatapa
NB!1-2-3-1-2-3-1-2
etc.etc........
Step-by-step FoleyManometer LV Instructions for Use
Initial set-up and preparations (use aseptic techniques):
Open the FoleyManometer pouch and close the tube clamp.
Place the urine collection device under the patient's bladder and tape the drainage tube to the bed sheet as shown on the drawing.
Insert the FoleyManometer between catheter and drainage device.
Prime the FoleyManometer with 20ml of sterile saline through its needle-free injection/sampling port. Prime only once i.e. at initial set-up, or subsequently to remove any air in the manometer tube.
Pvesical Monitoring
Place the "0 mmHg" mark of the manometer tube at the symph. pubis level and elevate the filter vertically above the patient.
Open clamp, and read Pvesical (end-expiration value) when the meniscus has stabilized.
Close clamp and place the FoleyManometer in its drainage position.
NB:
Prime only once i.e. at initial set-up, or subsequently to remove any air in the manometer tube which should always be fluid filled.
Slow descent (>20 - 30 seconds) of the meniscus during a Pvesical determination suggests a blocked or kinked Foley catheter.
Urine sampling from the needle-free port is facilitated by temporarily opening the red clamp. Remember to close clamp afterwards.
Avoid a U-bend of the large urimeter drainage tube (which will impede urine drainage).
Replace the FoleyManometer whenever the Foley catheter or the urine collection device is replaced. Do not use the FoleyManometer for more than 7 days.